Highland Technology V420

From WikiProducts, the products encyclopedia

Jump to: navigation, search

Contents

Overview


The V420 is an 8-channel, isolated, programmable resistance simulator. It uses entirely solid-state simulation to eliminate transient errors associated with relay switching, and provide for monotonic resistance changes. Channels are individually programmable in four resistance ranges, from 5 ohms to 65.5K on all ranges. Channels can operate from microvolts up to 24V / 50mA. Overload protection and channel error detection are included.

Image:V420-2R.jpg

Any channel may be rerouted, under software control, to the front-panel calibration-check connector, allowing the board to be tested for calibration by a precision DVM without disturbing the connections to the equipment under test. The BIST option adds an onboard precision ohmmeter, allowing loopback checking and full automatic self-test. Optional external termination and bridge-completion boards are available.

Features

  • Eight channel, isolated, wide-range, 2/4 wire resistance simulator.
  • Fast, monotonic, glitch-free resistance value programming.
  • Suitable for simulating resistors, platinum/copper RTDs, thermistors, strain gages, load cells.
  • Users may directly program resistances or RTD simulation (copper or platinum) temperatures.
  • Four programmable resistance ranges allow resistances from 5 ohms to 65.5 Kohms.
  • Any channel is switchable to front-panel calibration check connector for in-system test without disconnecting field wiring.
  • Simple register based VME interface with no handshaking required.
  • Optional built-in self-test (BIST) and mil/COTS versions available.
  • Optional termination panels and bridge completion boards available.

Specifications

V420 VME RESISTANCE SIMULATOR MODULE
FUNCTION
8-channel isolated resistance/RTD simulator
COMPLIANCE
IEEE 1014 VME; A16:A24:D16; no byte writes.
Platinum RTD tables per IEC-751 for "385" curve RTDs.
Platinum RTD tables per ITC-90 for "393" curve RTDs.
PACKAGING
U single-wide VME module
CHANNEL INPUTS
8 channels, 2- or 4-wire resistance simulation, galvanically isolated
RANGES
Programmable per channel.
5 R - 500 R
50 R - 5000 R
500 R - 50 KR
5KR - 65KR
120 - +650 °C 100R/1K Pt385 RTD simulation
-120 - +900 °C 100R/1K ITC-90 RTD simulation
-100 - +400 °C 10R Copper RTD simulation
RESOLUTION
Resistance: Varies within each range: 15 bits at low end; 12 bits at high end.
RTD: 0.0625°C resolution
ACCURACY
Resistance +-0.1% 4-wire, DC or AC to 1 KHz; 1% accurate to 20 KHz. Settles to 0.1% within 50 microseconds.
RTD simulation is accurate to +-0.1 °C.
Guaranteed monotonic within each range.
Guaranteed glitch-free value changes within each range.
DC offset error < 10 microvolts
TEMPERATURE COEFFICIENT
Below 40 PPM/K; below 50 PPM/watt
MAXIMUM INPUT
Differential: +-50 mA, +-24 V
Common Mode: +-750 V
ESD to 15KV, human body model.
CONNECTORS
Two D-25 females, with four 4-wire channels each
One D9 calibration check connector
INDICATORS
LEDs indicate VME access, CPU activity, error conditions; additional LED is user programmable.
CALIBRATION INTERVAL
One year
OPERATING TEMP
0 to 60 ºC; extended MIL/COTS ranges available
VERSIONS
V420-1 0/60°C
V420-2 0/60°C with BIST
V420-3 -20/80°C MIL/COTS, coated, BIST


Manuals, drivers, downloads & other resources

http://www.highlandtechnology.com/DSS/V420DS.html


See also


V470 VME ANALOG OUTPUT MODULE

Dealers, trainers, support, maintenance & service centers

Highland Technology, Inc.

United States

California

San Francisco

P: (415) 551-1700

F: (415) 551-5129

http://www.highlandtechnology.com

info@highlandtechnology.com


Personal tools