Highland Technology T250
From WikiProducts, the products encyclopedia
Contents |
Overview
FAST IMPULSE GENERATOR
Optical trigger versions compatible with Highland optical transmitter products:
V880 System Timing Module
V720 Electrical/Optical Converter VME Module
J720 Electrical/Optical Converter
J724 Electrical/Optical Converter
V951-6 Optical-Output VME Delay Generator
Typical waveform driving 50-ohm load.
Features
- Generates 65 volt, 350 ps positive pulses.
- Compact, rugged extruded package allows the pulse generator to be located wherever most convenient, including embedded applications.
- Accepts trigger inputs from the Highland electrical signal generators or compatible sources. Optionally equipped units can also accept optical triggers from Highand e/o converters or compatible sources.
- Adjustable logic-level inputs are compatible with TTL, LVTTL, ECL, NIM and most other common logic levels.
- Optionally equipped optical trigger version T250-1 uses commonly available 62/125-m, 850-nm, ST-connectorized multimode fiber with user-adjustable optical threshold level. The T250-3 & T250-5 versions are suited for use with singlemode or multimode fiber at 1310 and 1550 nm.
- On-board O/E converter provides fast analog electrical output for monitoring fiberoptic link quality.
- Timing jitter typically below 3 ps RMS.
- Ideal accessory for Highland Model P400 Digital Delay/Pulse Generator
Specifications
| FUNCTION | Single-channel fast electrical impulse generator |
| PACKAGING | 4.0 x 4.75 x 1.25 inches extruded aluminum enclosure. |
| TRIGGER | T250-9 Logic level; SMB connector standard. Threshold adjustable from -4.0 to +5.0 volts range. Selectable 50 ohms / HiZ termination. Rising edge triggers pulse; internal switch allows polarity inversion. T250-1 Adds ST-connectorized optical 850 nm nom. wavelength trigger detector with user-adjustable threshold. T250-3 Adds ST-connectorized optical 1310 nm nom. wavelength trigger detector with user-adjustable threshold. T250-5 Adds ST-connectorized optical 1550 nm nom. wavelength trigger detector with user-adjustable threshold. Optical Threshold adjustable from 100 uW to 800 uW. Factory threshold setting 300 uW |
| ANALOG O/E OUTPUT | Optically triggered models feature an internal O/E converter that generates +1 volt per mW optical input into 50-ohm load. Source impedance 50 ohms, Bandwidth 250 MHz nom., DC-coupled, SMB jack connector |
| PULSE OUTPUT AMPLITUDE | +65 +/- 2 volts peak into 50 ohms, < 100 mV/deg-C |
| PULSE WIDTH | 300-350 ps FWHM |
| RISE TIME | < 300 ps, 10% - 90% |
| FALL TIME | < 300 ps, 10% - 90% |
| PRECURSOR and RINGDOWN | < 12 % |
| REPETITION RATE | 0 - 1 MHz |
| COUPLING | Digital: DC, rising-edge triggered |
| PROPAGATION DELAY | 10.0 +/- 500 ps from electrical trigger |
| TEMPERATURE COEFFICIENT | Below 2 ps/deg C |
| POWER | +15 Volts at 400 mA nom. 2.1 x 5 mm connector. Optional high retention-force MMCX connector |
| JITTER | Below 3 ps RMS |
| INDICATORS | Front panel green power LED, blue trigger LED |
| ENVIRONMENT | 0 to 70 degrees C, operating. 0 to 95% humidity, non-condensing. |
| COMPLIANCE | OEM product has no UL/FCC/CE compliance requirements. Designed to meet UL/FCC/CE requirements. Power adapter when furnished is UL/CE certified. |
| ACCESSORIES | J15 15 volt wall-plug power supply, USA plug. J14 International plug adapter Kit. J53 SMB-BNC Cable, 36-in length. J732 Mounting flange. J720 Electrical-to-Optical converter. |
Manuals, drivers, downloads & other resources
http://www.highlandtechnology.com/DSS/T250DS.html
See also
P400 Benchtop digital delay generator and pulse generator
T560 Embedded small-size delay generator
T564 Embedded delay and pulse train generator
T200 Bipolar Impulse generator
V350 Function Generator
V850/851 4- and 6-channel VME digital delay generators
V880 VME precision timing and pulse generator module
V951 VXI digital delay generator
Dealers, trainers, support, maintenance & service centers
Highland Technology, Inc.
United States
California
San Francisco
P: (415) 551-1700
F: (415) 551-5129
http://www.highlandtechnology.com
info@highlandtechnology.com